HOW TO USE STEP GAUGE ?

HOW TO USE STEP GAUGE ?

1. Notes

    Step Gauges (made of aligned precision blocks) are precision measuring instruments. Abnormal force must be prevented during operation. Ill practices include, lifting the step gauge by grabbing the blocks, striking the blocks with any objects, bringing abnormal force to blocks, dropping the step gauge from a height, bending the body of the step gauge with powerful force and etc.


     Anyi Step Gauge has two kinds of blocks: ceramic and steel. Attention must be paid to the rust protection of steel blocks and the spacers between blocks. Under harsh environment, attention must also be paid to the rust protection of the stainless steel body of Anyi Step Gauge.


Go to the Step Gauge page


2. Preparation before use


    1). Take out the Step Gauge from the box carefully (please refer to the above notes). Dip the cleaning absorbent cotton into some recommended cleaner, like gasoline #120, and clean the working surfaces of Step Gauge and the surface of the base (avoid dissolving detergent).

 
    2).When calibrating the positioning accuracy of X-axis or Y-axis of CNC machines, the body of  step gauge must be placed parallel to the guide rail (the obliquity 100: 0.01). When calibrating the Z-axis, stand the Step Gauge vertically on the supporting surface. If the perpendicularity of the supporting surface and Z-axis is not good, it is important to make correction. When fixing the step gauge, the clamping points must be the supporting parts of the Airy points of the Step Gauge. Clamping mustn't cause bending and deforming.


    3) Fix a micrometer dial on the suitable place of the equipment to be calibrated. This micrometer dial can be a micron dial test indicator or a horizontal stylus probe. Expected accuracy decides the type of micrometer dial to be used. The repeatability must be as good as possible. When calibrating Coordinate Measuring Machines (CMM) and Height Gauges, we can use their own probes for the calibration without the presence of a micrometer dial . When clamping the micrometer dial with a magnetic stand, the rigidity and stability of the magnetic stand must be good.


      a) For Machining Centers and other similar machine tools, the micrometer dial must be fixed on the spindle head frame.
      b) For lathes, the micrometer dial must be fixed on the cutter frame.
      c) For others, the fixing place is similar to the above.

    4). Same Temperature: After adjusting the positioning of step gauge and mirometer dial, wait 3-4hrs until the step gauge assumes the same as or very similar temperature to the object to be calibrated. The time of waiting depends on the temperature range of step gauge and the equipment to be calibrated and the ambient temperature. Under normal circustances, 4-8 hours is good. 


        To achieve higher accuracy, it is important to extend the time of waiting.  The time of waiting 12-24 hours. But again, this also depends on the temperature difference beteen the step gauge and the equipment to be calibratied. 


    5). Warm-up of equipment
    Ater the time of waiting for temperature matching, the equipment to be calibrated should be in sufficient operation before the calibration begins. At this time, the equipment's temperature and accurace, etc. shows its real performance , the step gauge is nearer the state of the object to be calibrated.


Go to the Step Gauge page


3. Calibrating
    

    1). Move the guide rail of the equipment to be calibrated to its origin, then make the stylus probe of the  micrometer dial touch the center of the origin working surface and compress the stylus probe, then set the enumerator of the guide rail and the micrometer dial to zero.


    2). Move the stylus probe along the direction which is perpendicular to the guide rail and out of the working surface of the gauge blocks.


    3). Move the stylus probe along the direction which is parallel to the guide rail and to the next point to be calibrated.


    4). Repeat the second procedure reversely, the displacement should be the same (namely move the stylus probe to the center of the current working surface of the gauge block.


    5). Fine adjustment of the equipment to be calibrated

    Two options: One is to adjust the equipment to the nominal value that the guide rail displays. Now the differential between the guide rail and the step gauge can be read directly from the micrometer dial. The other is to adjust the equipment to the origin of the micrometer dial, now the differential between the guide rail and the step gauge can be read directly from the enumerator of the guide rail. The first method is the first choice when the repeatability, resolution and accuracy of the micrometer dial are higher than those of the equipment to be calibrated, this method gives precision measuring. One measurement can use only one method. Two methods can't be used in one measurement.


  Special notice should be taken to the parts to be calibrated: The step gauge has uni-directional and bi-directional working surfaces. The working surfaces which are in the same direction that the guide rail moves are usually used. If the uni-directional and bi-directional working surfaces along a direction are used, the error of reverse of direction of the guide rail and the micrometer dial will be included in the calibration result, unless the error is very small or the error is the calibration target.


Go to the Step Gauge page


4. Data processing

    The measured value plus the error of the step gauge's relevant point are the error of the calibrated point of the guide rail. The symbol of plus and minus must be paid attention to when processing the data. For high precision calibration, the coefficient of thermal expansion, the temperature deviation from standard temperature and the error of step gauge, etc. should be considered in data processing. 


    After data processing, correction to the positioning accuracy of the guide rail can be carried out based on the result.